EasyManuals Logo

Agilent Technologies 1260 Infinity Quick Start Guide

Agilent Technologies 1260 Infinity
98 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #81 background imageLoading...
Page #81 background image
Purification Solution - Developer's Guide 81
Calibration Procedures
6
System Configuration and Delay Time Calibration
Decrease the volume of tubing between the splitter and the detector(s) in
the make-up flow path.
After one or more solutions have been applied, repeat the fraction collector
delay calibration from the beginning.
Insufficient MSD-to-FC delay time
The minimum required delay time for MSD triggered collection either 5 sec or
6 × (MSD cycle time), whichever is the higher.
The reasons for insufficient MSD-to-FC delay time could be the following:
The volume of delay tubing in front of the FC is too low.
The MSD cycle time is too high.
The make-up flow is too low for the applied preparative flow. There are
limitations to the combinations of preparative and make-up flows; see the
example in “System Configuration and Delay Time Calibration” on page 70.
The volume of tubing between the splitter and detector(s) is too high.
The following suggestions are options for increasing the MSD-to-FC delay time,
but each of them has a certain disadvantage (given in parentheses) so
consider the most suitable way for the given situation:
Optimize the MSD cycle time in the MSD method (a cycle time higher than
0.83 sec will increase the required delay volume limit):
Use only one polarity if suitable (requires detailed knowledge of
collected compounds).
Optimize peakwidth parameter: decrease it until MSD signal noise is still
acceptable.
Decrease the MW range or increase the MW step size if suitable: typically,
it will not directly decrease the cycle time but it allows further
decreasing of the peakwidth parameter (a short MW range hides
potential contaminants).
Consider using the MSD (ultra)fast mode, which enables a lower cycle
time.
Performing MSD maintenance can also help in optimization of the noise
level.
Use a longer delay coil in front of the FC (a higher coil volume causes more
dispersion, so use coils with a reduced dispersion rate such as knitted coils;

Table of Contents

Other manuals for Agilent Technologies 1260 Infinity

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 Infinity and is the answer not in the manual?

Agilent Technologies 1260 Infinity Specifications

General IconGeneral
BrandAgilent Technologies
Model1260 Infinity
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals