82 Purification Solution - Developer's Guide
6
Calibration Procedures
System Configuration and Delay Time Calibration
also, a longer coil with the same inner diameter will increase
back-pressure).
Decrease the preparative flow (influences the chromatographic resolution).
Increase the make-up flow (increases speed of the analytes when passing
through detectors, which typically results in narrower and/or smaller
peaks, so consider whether or not the detector data rate is sufficient for the
signal response).
Decrease the volume of tubing between the splitter and the detector(s).
After one or more solutions have been applied, repeat the fraction collector
delay calibration from the beginning.