EasyManuals Logo

Agilent Technologies 1260 Infinity Quick Start Guide

Agilent Technologies 1260 Infinity
98 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #84 background imageLoading...
Page #84 background image
84 Purification Solution - Developer's Guide
7
Checkout Procedure
System Configuration and Delay Time Calibration
10 mm UV cell (1 μL injection): Dilute 20× (1 mL = 50 μL sample + 950 μL
solvent)
60 mm UV cell (1 μL injection): Dilute 100× (1 mL = 10 μL sample +
990 μL solvent)
Plug vial solvent: 2 % acetonitrile (or Solvent A or other compatible
solvent). Refresh the solvent frequently to avoid bacterial growth and
injection cross-contamination.
Recommended: For best column performance and lifetime, filter the sample
in DMSO solvent before use with the PTFE syringe filter (5190-5084). Filter
other sample solvents (water, acetonitrile) such as the plug solvent with the
regenerated cellulose syringe filter (5190-5108). To select the most suitable
filter type with the highest sample recovery, filter the sample solvent matrix
without the sample first.

Table of Contents

Other manuals for Agilent Technologies 1260 Infinity

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 Infinity and is the answer not in the manual?

Agilent Technologies 1260 Infinity Specifications

General IconGeneral
BrandAgilent Technologies
Model1260 Infinity
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals