Purification Solution - Developer's Guide 67
Calibration Procedures
6
Characterizing the column volume
Save the Method as
Column_Volume.
Set the Stoptime to No Limit for all modules (infinite run time).
Pump:
Set Solvents: 75 % B.
Clear the TimeTable.
Injection volume 2 μL (important: increase the volume if the marker
signal is too low).
If you are using the dual loop autosampler:
Select Lower injection loop.
Select Partial loop filling.
Use Plugs settings as described in “Analytical Method Settings” on
page 16 of the Default Purification Method.
UV detector:
Thiourea: Set the Signal A wavelength to 242 nm with 4 nm bandwidth
and no reference.
Uracil: Set the Signal A wavelength to 260 nm with 4 nm bandwidth
and no reference.
Save the method changes.
6 Display the relevant UV wavelength profile (242 nm or 260 nm) in the Online
Plot.
7 If you are using the dual loop autosampler:
equilibrate the upper loop:
Switch the valve to the upper loop
Right-click on the autosampler diagram and click Switch Valve to
Upper Loop, If the command is not available, the upper loop is
already selected.
Apply a pump flow of 4 mL/min for 3 min.
Switch the valve to the lower loop.
Wash the needle:
right-click on autosampler diagram
Wash Needle > Flush Port15 s
Purge the needle:

Table of Contents

Other manuals for Agilent Technologies 1260 Infinity

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 Infinity and is the answer not in the manual?

Agilent Technologies 1260 Infinity Specifications

General IconGeneral
BrandAgilent Technologies
Model1260 Infinity
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals