EasyManua.ls Logo

Agilent Technologies 1260 Infinity - How to Optimize the Compressibility Compensation Setting; Solvent Compressibility Calibration

Agilent Technologies 1260 Infinity
98 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Purification Solution - Developer's Guide 87
Checkout Procedure
7
Acquire an analytical run using the Chemstation sequence
• 60 mm UV cell: dilute 100× (1 mL = 10 μL sample + 990 μL solvent).
3 Place 1 mL filtered sample in the autosampler.
4 Create a new sequence: Sequence > New Sequence Template
5 Set Sequence > Sequence Parameters (for example, folder path, shutdown
macro).
6 Complete the sequence table: Sequence > Sequence Table:
• Location: sample position.
• Sample name: Check-out sample
• Method name: select the analytical method (Analytical_Gradient)
• Inj/Location: 1
• Leave Sample amount field empty
• Click OK
7 Sequence > Save Sequence Template AsCheckout_Sample
8 Display the 263 nm UV profile (signal A) in the Online Plot.
9 Turn on the pump.
10 For the dual-loop autosampler:
a Equilibrate the upper loop:
a Switch the valve to the upper loop (right-click on the autosampler
diagram and select Switch Valve to Upper Loop; if the command is not
available, the upper loop is already selected).
b Apply a flow of 2mL/min for 5min
c Switch the valve to the lower loop
b Wash the needle (right-click on the autosampler diagram and select Wash
Needle > Flush Port > 15 s)
c Purge the needle (right-click on the autosampler diagram and select Start
Purging > 3×)
11 Set the target analytical flow (default is 1 mL/min for a 4.6 mm ID column
and 0.6 mL/min for a 2.1 mm ID column).
12 Equilibrate the column until the pressure and UV signal are stable.
13 RunControl > Run Sequence.
14 Rerun the sample at least once and check the results of the sequence
visually (peak shape, resolution). If the UV absorbance of five major peaks

Table of Contents