100 1260 Infinity II Analytical-Scale & Bio-inert Fraction Collector User Manual
6
Error Information
Module Error Messages
Module Error Messages
Exhaust Fan Failed
Error ID: 4456, 4457
The exhaust fan in the module has failed.
The hall sensor on the fan shaft is used by the main board to monitor the fan
speed. If the fan speed falls below a certain value the error message is
generated and the module shuts down.
Front Door Error
Error ID: 4750
The front door and/or the SLF board are damaged.
Probable cause Suggested actions
1
Fan cable disconnected.
Please contact your Agilent service
representative.
2
Defective fan.
Please contact your Agilent service
representative.
3
Defective main board.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1
The flat ribbon cable from MTP main board
to the SLF board is not connected correctly.
Please contact your Agilent service
representative.
2
The sensor on the SLF board is defective.
Please contact your Agilent service
representative.
3
The door is bent or the magnet is
misplaced/broken.
Change the side door.
4
Defective MTP main board.
Please contact your Agilent service
representative.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260Infinity II Bio-inert FractionCollector and is the answer not in the manual?

Agilent Technologies 1260Infinity II Bio-inert FractionCollector Specifications

General IconGeneral
BrandAgilent Technologies
Model1260Infinity II Bio-inert FractionCollector
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals