160 1260 Infinity II Analytical-Scale & Bio-inert Fraction Collector User Manual
8
Parts for Maintenance and Repair
List of Recommended Plates and Closing Mats
List of Recommended Plates and Closing Mats
Warnings
WARNING
Explosive gas mixtures
There is a risk of explosive gas mixtures in the instrument if flammable solvents are
used.
Cover the plates.
Remove the plates from the fraction collector after turning it 0FF.
Only use solvents with a flash point higher than 200 °C.
CAUTION
Contamination with adhesives
Closing mats with adhesives can give some contamination in the system. The
adhesive is soluble in most of the solvents used in HPLC.
In general do not use closing mats with adhesive. The fraction collector has no
prepunch needle, therefore the adhesive will clog the needle after several
injections.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260Infinity II Bio-inert FractionCollector and is the answer not in the manual?

Agilent Technologies 1260Infinity II Bio-inert FractionCollector Specifications

General IconGeneral
BrandAgilent Technologies
Model1260Infinity II Bio-inert FractionCollector
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals