1260 Infinity II Analytical-Scale & Bio-inert Fraction Collector User Manual 125
Maintenance
7
Exchange the Analytical Needle Assembly
Exchange the Analytical Needle Assembly
For bio-inert modules use bio-inert parts only!
When When the needle is visibly damaged
When the needle is blocked or contaminated
Tools required p/n Description
8710-1534 Wrench, open end, 4 mm
8710-0510 Wrench open 1/4 — 5/16 inch
Parts required # p/n Description
1 G1367-87200 Analytical needle assembly
1 G5667-87200 Needle assembly (bio-inert)
Preparations Position the transport unit of the fraction collector in the Home position.
Remove all installed trays from the tray base.
Position the transport unit of the fraction collector in the Change Parts position and turn off the
instrument.
Remove the needle from its carrier before unscrewing the needle tubing.
WARNING
Personal injury
Risk of personal injury caused by the needle arm movement.
Keep fingers away from the needle area.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260Infinity II Bio-inert FractionCollector and is the answer not in the manual?

Agilent Technologies 1260Infinity II Bio-inert FractionCollector Specifications

General IconGeneral
BrandAgilent Technologies
Model1260Infinity II Bio-inert FractionCollector
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals