1260 Infinity II Analytical-Scale & Bio-inert Fraction Collector User Manual 23
Site Requirements and Specifications
2
Physical Specifications
Physical Specifications
Table 2 Physical Specifications
Type Specification Comments
Weight 13.5 kg (29.8 lbs)
w/o Thermostat
Dimensions
(height × width × depth)
200 x 345 × 440 mm
(8 x 13.5 × 17 inches)
Line voltage 100 – 240 V~, ± 10 % Wide-ranging
capability
Line frequency 50 or 60 Hz, ± 5 %
Power consumption 200 VA / 180 W
Ambient operating
temperature
4 – 40 °C (41 – 104 °F)
Ambient non-operating
temperature
-40 – 70 °C (-40 – 158 °F)
Humidity
< 95 %, at 25 – 40 °C (77 – 104 °F)
1
1
If a thermostat is used the upper value for humidity can be reduced. Please check your lab condi-
tions to stay beyond dew point values for non–condensing operation.
Non-condensing
Operating altitude Up to 3000 m (9842 ft)
Non-operating altitude Up to 4600 m (15092 ft) For storing the module
Safety standards:
IEC, EN, CSA, UL
Installation category II, Pollution degree 2 For indoor use only.
ISM Classification ISM Group 1 Class B According to CISPR 11
Permitted solvents Boiling point ≥56 °C
Auto-ignition temperature ≥200 °C
Ignition Class IIA, IIB (IEC60079-20-1)