Error Information
Module Error Messages
Adapter Required
Error ID: 4961
The wellplate adapter is required for the following operation.
Funnel Not Supported
Error ID: 4962
Funnels are only supported for the analytical/bio-inert fraction collector.
Consequently the error message is displayed, if a tray with funnels is
configured for the preparative fraction collector.
Probable cause Suggested actions
1
Wellplate adapter not attached.
Attach wellplate adapter.
2
Cable between needle carrier assembly to
transport unit assembly disconnected or
defective.
Check cable between needle carrier assembly
to transport unit assembly.
3
Needle carrier assembly defective.
Exchange needle carrier assembly.
Probable cause Suggested actions
1
Fraction collector type and tray not
compatible.
• Check the tray configuration in your CDS.
Funnel trays are only supported for the
analytical scale fraction collector.
• Check the system configuration in your
CDS, and verify that an analytical scale
fraction collector is configured.