108 1260 Infinity II Analytical-Scale & Bio-inert Fraction Collector User Manual
6
Error Information
Module Error Messages
Initialization Failed
Error ID: 4950
The fraction collector failed to complete initialization correctly. The fraction
collector initialization procedure moves the needle arm and transport
assembly to their home positions in a predefined routine. During
initialization, the processor monitors the position sensors and motor encoders
to check for correct movement. If one or more of the movements is not
successful, or is not detected, the error message is generated.
Probable cause Suggested actions
1
Transport unit not aligned correctly
Perform an auto-alignment.
2
Mechanical obstruction.
Ensure unobstructed movement of the
transport assembly.
3
Defective transport assembly flex board.
Please contact your Agilent service
representative.
4
Defective MTP main board.
Please contact your Agilent service
representative.
5
Excessive weight on top of the fraction
collector (see also “Side Door Error” on
page 101 )
Check stack configuration and reduce weight
on top of the fraction collector.