1260 Infinity II Analytical-Scale & Bio-inert Fraction Collector User Manual 107
Error Information
6
Module Error Messages
Pusher Wrong or Defect
Error ID: 4965
Wrong or Missing Needle (Analytical Scale)
Error ID: 4966
Probable cause Suggested actions
1
Wellplate adapter not attached.
Attach wellplate adapter.
2
Needle carrier assembly defective.
Exchange needle carrier assembly.
Probable cause Suggested actions
1
No needle installed.
Check which needle has been installed.
2
Short needle for high flow rates installed,
but the wellplate adapter hasn’t been
removed.
Remove wellplate adapter if short needle for
semi-preparative operation is installed.