96 1260 Infinity II Analytical-Scale & Bio-inert Fraction Collector User Manual
6
Error Information
General Error Messages
Leak Sensor Short
Error ID: 0082
The leak sensor in the module has failed (short circuit).
The current through the leak sensor is dependent on temperature. A leak is
detected when solvent cools the leak sensor, causing the leak sensor current to
change within defined limits. If the current increases above the upper limit,
the error message is generated.
Leak Sensor Open
Error ID: 0083
The leak sensor in the module has failed (open circuit).
The current through the leak sensor is dependent on temperature. A leak is
detected when solvent cools the leak sensor, causing the leak-sensor current to
change within defined limits. If the current falls outside the lower limit, the
error message is generated.
Probable cause Suggested actions
1
Defective leak sensor.
Please contact your Agilent service
representative.
2
Leak sensor incorrectly routed, being
pinched by a metal component.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1
Leak sensor not connected to the main
board.
Please contact your Agilent service
representative.
2
Defective leak sensor.
Please contact your Agilent service
representative.
3
Leak sensor incorrectly routed, being
pinched by a metal component.
Please contact your Agilent service
representative.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260Infinity II Bio-inert FractionCollector and is the answer not in the manual?

Agilent Technologies 1260Infinity II Bio-inert FractionCollector Specifications

General IconGeneral
BrandAgilent Technologies
Model1260Infinity II Bio-inert FractionCollector
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals