110 1260 Infinity II Analytical-Scale & Bio-inert Fraction Collector User Manual
6
Error Information
Module Error Messages
Movement to Next Position Failed
Error ID: 4957
The transport mechanism detected an unexpected situation during the
movement to the next fraction position.
Could Not Find a Valid Next Position
Error ID: 4958
There has been more fractions than fraction positions.
Probable cause Suggested actions
1
Mismatch between tray configuration and
the loading of the trays, e.g. 4 well plates
are configured in the UI, but only three are
loaded, or shallow plates are configured,
but deep well-plates are used, or tube
height doesn't match the configuration.
Check the configuration and the loading of the
tray.
Probable cause Suggested actions
1
Unexpected number of fractions.
• If possible use a tray with more positions.
• Add an additional fraction collector to the
system.
2
Start of an analysis / sequence / run
without changing of the tray of the previous
run.
Ensure that there are enough fraction positions
for the complete analysis / sequence / run.