118 1260 Infinity II Analytical-Scale & Bio-inert Fraction Collector User Manual
7
Maintenance
Replace the Inlet/Waste Tubings (Analytical and Bio-inert)
CAUTION
Liquid spills or fraction losses
Worn or damaged tubings can cause potential spills or lead to fraction losses.
Explicitly follow the described installation procedures to maximize the lifetime of
inlet / waste tubing assembly and the valve to needle tubing, and to avoid potential
spills or fraction losses.
Regularly inspect the tubings and exchange them if they are worn out or show
visible signs of damage.
NOTE
Connect the tubings as described in order to maximize their lifetime and operating
security.
1
Locate the diverter valve with the finger tight fittings of
the inner tubing kit.
2
Unscrew the two finger-tight fittings of the inner tubing
kit at the diverter valve.
3
Unclip the inlet / waste tubing assembly from the leak
adapter.
4
Remove the outer tubing kit from the unions.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260Infinity II Bio-inert FractionCollector and is the answer not in the manual?

Agilent Technologies 1260Infinity II Bio-inert FractionCollector Specifications

General IconGeneral
BrandAgilent Technologies
Model1260Infinity II Bio-inert FractionCollector
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals