Preparing the Fraction Collector
4
Solvent Information
Diamond-Like Carbon (DLC)
Diamond-Like Carbon is inert to almost all common acids, bases and solvents.
There are no documented incompatibilities for HPLC applications.
Fused silica and Quartz (SiO
2
)
Fused silica is used in Max Light Cartridges. Quartz is used for classical flow
cell windows. It is inert against all common solvents and acids except
hydrofluoric acid and acidic solvents containing fluorides. It is corroded by
strong bases and should not be used above pH 12 at room temperature. The
corrosion of flow cell windows can negatively affect measurement results. For
a pH greater than 12, the use of flow cells with sapphire windows is
recommended.
Gold
Gold is inert to all common HPLC solvents, acids and bases within the
specified pH range. It can be corroded by complexing cyanides and
concentrated acids like aqua regia.
Zirconium Oxide (ZrO
2
)
Zirconium Oxide is inert to almost all common acids, bases and solvents.
There are no documented incompatibilities for HPLC applications.
Platinum/Iridium
Platinum/Iridium is inert to almost all common acids, bases and solvents.
There are no documented incompatibilities for HPLC applications.
Fluorinated polymers (PTFE, PFA, FEP, FFKM, PVDF)
Fluorinated polymers like PTFE (polytetrafluorethylene), PFA
(perfluoroalkoxy) and FEP (fluorinated ethylene propylene) are inert to
almost all common acids, bases, and solvents. FFKM is perfluorinated rubber,
which is also resistant to most chemicals. As an elastomer, it may swell in
some organic solvents like halogenated hydrocarbons.
TFE/PDD copolymer tubings, which are used in all Agilent degassers except
1322A, are not compatible with fluorinated solvents like Freon, Fluorinert, or
Vertrel. They have limited life time in the presence of Hexafluoroisopropanol

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260Infinity II Bio-inert FractionCollector and is the answer not in the manual?

Agilent Technologies 1260Infinity II Bio-inert FractionCollector Specifications

General IconGeneral
BrandAgilent Technologies
Model1260Infinity II Bio-inert FractionCollector
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals