1260 Infinity II Analytical-Scale & Bio-inert Fraction Collector User Manual 71
Preparing the Fraction Collector
4
Install Capillary Connections
3
Carefully slide the ferrule components on after the nut
and then finger-tighten the assembly while ensuring that
the tubing is completely seated in the bottom of the end
fitting.
4
Use a column or injection valve to gently tighten the
fitting which forces the ferrule to seat onto the tubing.
NOTE
Don't overtighten. Overtightening will shorten the
lifetime of the fitting.