1260 Infinity II Analytical-Scale & Bio-inert Fraction Collector User Manual 79
Preparing the Fraction Collector
4
Method Parameter Settings
Table 1 1 Method Parameter Settings
Parameter Description
Collection Behavior
Use this setting to either enable or disable the
fraction collection parameters of the instrument.
Peak Triggers
Use the Peak Triggers table to specify the
detection settings of the peak detectors available
in your system.
Peak Detection Mode
The following detection modes are available:
Off (The peak detector is not used)
Slope (Detects peaks based on slope
values only)
Limits: Up slope: 0.01 – 10000 units/s,
Down slope: 0.01 – 10000 units/s
Threshold (Detects peaks based on
threshold values only)
Limits: Threshold: -10000 –
10000 units,Upper threshold: 0.01 –
10000 units
Threshold and Slope (Detects peaks
based on both threshold and slope values)
Max Peak Duration
You can Limit Peak Duration to stop the
fraction collection in cases where the
baseline drifts and the signal does not
drop below the specified threshold value.
Limits: 1 – 10000 s

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260Infinity II Bio-inert FractionCollector and is the answer not in the manual?

Agilent Technologies 1260Infinity II Bio-inert FractionCollector Specifications

General IconGeneral
BrandAgilent Technologies
Model1260Infinity II Bio-inert FractionCollector
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals