3-60 Agilent 4155B/4156B Users Guide Vol.2, Edition 5
Measurement Functions
Compliance
Table 3-9 I Compliance Resolution
HPSMU 2 V 1 pA to 1000 mA
20 V 1 pA to 1000 mA
40 V 1 pA to 500 mA
100 V 1 pA to 125 mA
200 V 1 pA to 50 mA
Unit I Compliance Resolution
HRSMU 100 fA £ | I | £ 100 pA 10 fA
100 pA < | I | £ 1 nA 100 fA
1 nA < | I | £ 10 nA 1 pA
10 nA < | I | £ 100 nA 10 pA
100 nA < | I | £ 1 mA 100 pA
1 mA < | I | £ 10 mA1 nA
10 mA < | I | £ 100 mA 10 nA
100 mA < | I | £ 1 mA 100 nA
1 mA < | I | £ 10 mA 1 mA
10 mA < | I | £ 100 mA 10 mA
MPSMU 1 pA £ | I | £ 1 nA 100 fA
1 nA < | I | £ 10 nA 1 pA
10 nA < | I | £ 100 nA 10 pA
100 nA < | I | £ 1 mA 100 pA
1 mA < | I | £ 10 mA1 nA
10 mA < | I | £ 100 mA 10 nA
100 mA < | I | £ 1 mA 100 nA
1 mA < | I | £ 10 mA 1 mA
10 mA < | I | £ 100 mA 10 mA
Unit Output Range I Compliance Setting Range