4263B
[:SENSe]:CORRection:COLLect:METHod
f
REFL2
j
REFL3
g
Sets
or queries
the
measurement
error
correction
method.
REFL2
OPEN/SHORT correction
REFL3
OPEN/SHORT/LOAD
correction
Y
ou must
take
the
required
correction
data
using
the
[:SENSe]:CORRection:COLLect[:ACQuire]
command.
F
or
details
,
see
\T
o
P
erform
Correction
"
in
Chapter
4
and \Correction
Functions of
the 4263B
"
in
Chapter
7.
*
[:SENSe]:CORRection:COLLect:METHod
is
set
to
REFL2
by
3
RST
.
Query
response
is
REFL2
or
REFL3
.
[:SENSe]:CORRection:D
A
T
A?
ST
ANdard
f
1
j
2
j
3
g
Returns
the
correction
data.
(Query
Only)
ST
ANdard1
OPEN
correction
data
ST
ANdard2
SHORT
correction
data
ST
ANdard3
LO
AD
correction
data
Query
response
is
two
numeric
values
(separamed
by
a
comma) in
<
NR3
>
format.
[:SENSe]:CORRection[:ST
A
T
e]
f
ON
j
OFF
j
1
j
0
g
Sets
or
queries
if the
measurement error
correction
function
is
enabled.
OFF
or
0
Disables
the
correction function.
ON
or
1
Enables
the
correction
function.
Query
response
is
0
or
1
.
[:SENSe]:FIMP
edance:APERture
<
numeric
value
>
[MS
j
S]
Sets
or queries
measurement
time
mode:
Short,
Medium,
or
Long.
<
numeric
value
>
is
,
numeric
0.025
(Short),
0.065
(Medium),
or
0.500
(Long)
(s)
Query
response
is
a
numeric
value
in
<
NR3
>
format.
[:SENSe]:FIMP
edance:CONT
act:VERify
f
ON
j
OFF
j
1
j
0
g
Sets
or
queries
if
the
contact
check function
is
enabled.
OFF
or
0
disables the contact check function.
ON
or
1
enables the contact check function.
The reference data for the contact check is taken by the OPEN/SHORT correction.
Query response is
0
or
1
.
[:SENSe]:FIMPedance:RANGe:AUTO
f
ON
j
OFF
j
1
j
0
g
Sets or queries if the auto range measurement mode is enabled.
OFF
or
0
Hold range mode
ON
or
1
Auto range mode
Query response is
0
or
1
.
5-26 GPIB Reference