126 5975 Series MSD Operation Manual for MassHunter
4 Operating in Chemical Ionization (CI) Mode
To Perform an NCI Autotune (Methane Reagent Gas)
Procedure
1 From the Tune and Vacuum Control view, load NCICH4.U (or an existing
tune file for the reagent gas you are using).
2 From the Setup menu select the CI Tune Wizard and follow the system
prompts.
Accept the default temperature and other settings.
If you use an existing tune file, be sure to save it with a new name if you
don’t want to overwrite the existing values.
3 Under the Tune menu, click CI Autotune.
There are no tune performance criteria. If autotune completes, it passes (See
Figure 24 on page 127). If the tune sets the electron multiplier voltage
(EMVolts) at or above 2600 V, however, you may not be able to acquire data
successfully if your method sets EMVolts to “+400” or higher.
CAUTION
Always verify MSD performance in EI before switching to CI operation. See “To Verify
System Performance” on page 79. Always set up the CI MSD in PCI with methane as
the reagent gas first, even if you are going to be using a different reagent gas or going
to run NCI.
CAUTION
Avoid tuning unless absolutely necessary; this will minimize PFDTD background noise
and help prevent ion source contamination.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5975 and is the answer not in the manual?

Agilent Technologies 5975 Specifications

General IconGeneral
BrandAgilent Technologies
Model5975
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals