152 5975 Series MSD Operation Manual for MassHunter
5 General Maintenance
To Replace a Filament in an EI Source
Materials needed
Filament assembly (G2590-60053)
Gloves, clean, lint-free
Large (8650-0030)
Small (8650-0029)
Hex ball driver, 1.5-mm (8710-1570)
Procedure
1 Vent the MSD. See “Venting the MSD” on page 65.
2 Open the analyzer chamber. See “To Open the Analyzer Chamber” on
page 88.
3 Remove the ion source. See “To Remove the EI Ion Source” on page 142.
4 Remove the gold plated screw and washer for the filament(s).
WARNING
The analyzer operates at high temperatures. Do not touch any part until you are sure
it is cool.