Operating in Electron Impact (EI) Mode 3
5975 Series MSD Operation Manual for MassHunter 57
LCP Menus
To access a particular menu option, press [Menu] until the desired menu
appears, then press [Item] until the desired menu item appears. Table 6
through Table 11 list the menus and selections.
NOTE
Many menu items, especially on the ChemStation, MS Parameters, and Maintenance
menus, have no effect when the instrument is acquiring data.
Table 6 ChemStation menu
Action Description
Run Method Displays the current method name and starts an analysis.
Run Sequence Displays the current sequence and starts a sequence.
Run Current Tune Displays the current tune file and starts an autotune (EI mode
only; CI tune must be started from the MassHunter Data
Acquisition).
# of Messages Displays the number of messages and the text of the most recent
message. Use the arrow keys to scroll through previous messages
(up to 20).
Release ChemStation Disassociates the MassHunter Data Acquisition from the MSD.
Connection Status Displays the LAN connection status for the MSD.
Remote = connected to MassHunter Data Acquisition online
session
Local = not connected to MassHunter Data Acquisition online
session
Name of Instrument Displays the name of the instrument if connected to MassHunter
Data Acquisition online session. The name of the instrument is
the name assigned to the MSD by the MassHunter Data
Acquisition Configuration dialogue.