EasyManuals Logo

Agilent Technologies 5975 User Manual

Agilent Technologies 5975
192 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #129 background imageLoading...
Page #129 background image
Operating in Chemical Ionization (CI) Mode 4
5975 Series MSD Operation Manual for MassHunter 129
To Verify NCI Performance
This procedure is for EI/PCI/NCI MSDs only.
Materials needed
Octafluoronaphthalene (OFN), 100 fg/µL (5188-5347)
Procedure
1 Verify that the MSD performs correctly in EI mode.
2 Load the NCICH4.U tune file, and accept the temperature setpoints.
3 Select Gas A and set flow to 40%.
4 In Tune and Vacuum Control view, run CI Autotune. See “To Perform an
NCI Autotune (Methane Reagent Gas)” on page 126.
Note that there are no criteria for a “passing” Autotune in CI. If the
Autotune completes, it passes.
5 Run the NCI sensitivity method: OFN_NCI.M using 2 µL of 100 fg/µL OFN.
6 Verify that the system conforms to the published sensitivity specification.
Please see the Agilent Web site at www.agilent.com/chem for
specifications.
CAUTION
Always verify MSD performance in EI before switching to CI operation. See “To Verify
System Performance” on page 79. Always set up the CI MSD in PCI first, even if you are
going to run NCI.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5975 and is the answer not in the manual?

Agilent Technologies 5975 Specifications

General IconGeneral
BrandAgilent Technologies
Model5975
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals