EasyManua.ls Logo

Agilent Technologies 5975

Agilent Technologies 5975
192 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Installing GC Columns 2
5975 Series MSD Operation Manual for MassHunter 39
To Prepare a Capillary Column for Installation
Materials needed
Capillary column
Column cutter, ceramic (5181-8836) or diamond (5183-4620)
Ferrules
0.27-mm id, for 0.10-mm id columns (5062-3518)
0.37-mm id, for 0.20-mm id columns (5062-3516)
0.40-mm id, for 0.25-mm id columns (5181-3323)
0.5-mm id, for 0.32-mm id columns (5062-3514)
0.8-mm id, for 0.53-mm id columns (5062-3512)
Gloves, clean
Large (8650-0030)
Small (8650-0029)
Inlet column nut (5181-8830 for Agilent 7890A, 7820A and 6890, or
5183-4732 for 6850)
Magnifying loupe
Septum (may be old, used inlet septum)
Procedure
1 Slide a septum, column nut, and conditioned ferrule onto the free end of the
column (Figure 7). The tapered end of the ferrule should point away from
the column nut.

Table of Contents

Related product manuals