EasyManuals Logo

Agilent Technologies 5975 User Manual

Agilent Technologies 5975
192 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #140 background imageLoading...
Page #140 background image
140 5975 Series MSD Operation Manual for MassHunter
5 General Maintenance
To Replace the Primary Fuses
Materials needed
Fuse, T8 A, 250 V (2110-0969) – 2 required
Screwdriver, flat-blade (8730-0002)
The most likely cause of failure of the primary fuses is a problem with the
foreline pump. If the primary fuses in your MSD fail, check the foreline pump.
Procedure
1 Vent the MSD and unplug the power cord from the electrical outlet.
If one of the primary fuses has failed, the MSD will already be off, but for
safety you should switch off the MSD and unplug the power cord. It is not
necessary to allow air into the analyzer chamber.
2 Turn one of the fuse holders (Figure 25) counterclockwise until it pops out.
The fuse holders are spring loaded.
3 Remove the old fuse from the fuse holder.
4 Install a new fuse in the fuse holder.
5 Reinstall the fuse holder.
WARNING
Never replace the primary fuses while the MSD is connected to a power source.
WARNING
If you are using hydrogen as a GC carrier gas, a power failure may allow it to
accumulate in the analyzer chamber. In that case, further precautions are required.
See “Hydrogen Safety” on page 21.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5975 and is the answer not in the manual?

Agilent Technologies 5975 Specifications

General IconGeneral
BrandAgilent Technologies
Model5975
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals