EasyManuals Logo

Agilent Technologies 5975 User Manual

Agilent Technologies 5975
192 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #169 background imageLoading...
Page #169 background image
CI Maintenance 6
5975 Series MSD Operation Manual for MassHunter 169
To Clean the CI Source
Materials needed
Abrasive paper (5061-5896)
Alumina abrasive powder (8660-0791)
Aluminum foil, clean
Cloths, clean (05980-60051)
Cotton swabs (5080-5400)
Glass beakers, 500 mL
Gloves, clean, lint-free
Large (8650-0030)
Small (8650-0029)
Solvents
Acetone, reagent grade
Methanol, reagent grade
Methylene chloride, reagent grade
Ultrasonic bath
Preparation
1 Disassemble the ion source. See “To Disassemble the CI Source” on
page 167.
2 Collect the following parts to be cleaned for a CI source:
Repeller
Source body
Drawout plate
Drawout cylinder
Ion focus lens
Entrance lens
These are the parts that contact the sample or ion beam. The other parts
normally should not require cleaning.
3 Clean the parts as described in “To Clean the EI Source” on page 146.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5975 and is the answer not in the manual?

Agilent Technologies 5975 Specifications

General IconGeneral
BrandAgilent Technologies
Model5975
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals