56 5975 Series MSD Operation Manual for MassHunter
3 Operating in Electron Impact (EI) Mode
Loading OS
The operating system of the instrument controller is being initialized.
<method> Complete <timestamp>
The run and subsequent data processing are done. The same message appears
even if the run was terminated prematurely.
Method Loaded <method name>
Method parameters were sent to the MSD.
MS locked by <computer name>
MS parameters can only be changed from the MassHunter Data Acquisition.
Press Sideplate
A reminder during startup to press the MSD sideplate to ensure an adequate
vacuum seal.
Run: <method> Acquiring <datafile>
A run is in progress; data is being acquired to the designated data file.
To view system status during startup
1 The following messages are displayed on the LCP display during startup:
Press sideplate
Loading OS
Press sideplate
Loading MSD Firmware
2 Continue to press the sideplate of the MSD until the MSD Ready message
appears. This helps the instrument to pump down more quickly.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5975 and is the answer not in the manual?

Agilent Technologies 5975 Specifications

General IconGeneral
BrandAgilent Technologies
Model5975
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals