EasyManua.ls Logo

Agilent Technologies 5975

Agilent Technologies 5975
192 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CI Maintenance 6
5975 Series MSD Operation Manual for MassHunter 167
To Disassemble the CI Source
Materials needed
Gloves, clean, lint-free
Large (8650-0030)
Small (8650-0029)
Hex ball driver, 1.5 mm (8710-1570)
Hex ball driver, 2.0 mm (8710-1804)
Wrench, open-end, 10 mm (8710-2353)
Procedure
1 Remove the ion source. See “To Remove the CI Source” on page 165.
2 Remove the filaments. Refer to “To Replace a Filament in a CI Source” on
page 173.
3 Separate the repeller assembly from the source body. The repeller
assembly includes the source heater block assembly, repeller, and related
parts.
4 Remove the repeller and ceramic insulator and separate them.
5 Remove the setscrew for the lenses.
6 Pull the lens assembly out of the source body.
7 Remove the drawout cylinder and drawout plate from the source body.
8 Separate the ion focus lens, entrance lens and insulator.

Table of Contents

Related product manuals