16 5975 Series MSD Operation Manual for MassHunter
1 Introduction
The gauge is required for chemical ionization (CI) operation.
Table 3 5975 series MSD models and features
Model
Feature G3170A
G3175A
G3171A
G3176A
G3172A G3174A
High vacuum pump Diffusion Standard
turbo
Performance
turbo
Performance
turbo
Optimal He column flow mL/min 1 1 1 to 2 1 to 2
Maximum recommended gas flow
mL/min
*
* Total gas flow into the MSD: column flow plus reagent gas flow (if applicable).
1.5 2.0 4.0 4
Maximum gas flow, mL/min
Expect degradation of spectral performance and sensitivity.
2 2.4 6.5 6.5
Max column id 0.25 mm
(30 m)
0.32 mm
(30 m)
0.53 mm
(30 m)
0.53 mm
(30 m)
CI capability No No No Yes
DIP
capability (3rd party)
Direct insertion probe.
Yes Yes Yes Yes