EasyManuals Logo

Agilent Technologies 5975 User Manual

Agilent Technologies 5975
192 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #16 background imageLoading...
Page #16 background image
16 5975 Series MSD Operation Manual for MassHunter
1 Introduction
The gauge is required for chemical ionization (CI) operation.
Table 3 5975 series MSD models and features
Model
Feature G3170A
G3175A
G3171A
G3176A
G3172A G3174A
High vacuum pump Diffusion Standard
turbo
Performance
turbo
Performance
turbo
Optimal He column flow mL/min 1 1 1 to 2 1 to 2
Maximum recommended gas flow
mL/min
*
* Total gas flow into the MSD: column flow plus reagent gas flow (if applicable).
1.5 2.0 4.0 4
Maximum gas flow, mL/min
Expect degradation of spectral performance and sensitivity.
2 2.4 6.5 6.5
Max column id 0.25 mm
(30 m)
0.32 mm
(30 m)
0.53 mm
(30 m)
0.53 mm
(30 m)
CI capability No No No Yes
DIP
capability (3rd party)
Direct insertion probe.
Yes Yes Yes Yes

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5975 and is the answer not in the manual?

Agilent Technologies 5975 Specifications

General IconGeneral
BrandAgilent Technologies
Model5975
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals