Operating in EI Mode 4
5977B Series MSD Operating Manual 105
6 Click OK to close this dialog and start the tune. If the MSD temperatures are
not stable, you are prompted to wait or override the wait by clicking
Override.
7 Wait for the tune to complete, and generate the report.
8 To evaluate the tune results, select Evaluate Tune from the Checkout menu in
Instrument Control view.
To view history of tune results, in the Instrument Control view select
Checkout>View Previous Tunes....
To manually tune your MSD or to perform special autotunes, from the View
menu select Tune and Vacuum Control view. See the manuals and online help
provided with your Agilent MassHunter GC/MS Acquisition software for
additional information about tuning.