6 General Maintenance
160 5977B Series MSD Operating Manual
To Clean the EI HES
Materials needed
Abrasive paper (5061-5896)
Alumina abrasive powder (393706201)
Aluminum foil, clean
Cloths, clean (05980-60051)
Cotton swabs (5080-5400)
Glass beakers, 500 mL
Gloves, clean, lint-free
Large (8650-0030)
Small (8650-0029)
Solvents
Methanol (reagent-grade)
Methylene chloride (reagent-grade)
Acetone (reagent-grade)
Ultrasonic bath
Procedure
1 Disassemble the EI HES. (See “To Disassemble the EI HES” on page 157.)
2 Collect the following parts to be cleaned: (See Figure 50 on page 161.)
Filament mount
Source mount (does not get abrasively or ultrasonically cleaned)
Repeller
Source body
Extractor lens (5)
Post extractor lens 1 (4)
Post extractor lens 2 (3)
Ion focus lens (2)
Entrance lens (1)