7 CI Maintenance
200 5977B Series MSD Operating Manual
To Remove the EI HES Radiator
Materials needed
• Gloves, clean, lint-free
• Large (8650-0030)
• Small (8650-0029)
• Tweezers (8710-2460)
• T10 Torx screwdriver (5182-3466)
Procedure
1 Remove the EI HES. (See “To Remove the EI HES” on page 154.)
2 Disconnect the wires from the EI HES radiator. (See “To
Connect/Disconnect Wiring to the EI HES Radiator” on page 201.)
3 Use a T10 Torx screwdriver to loosen the 2 screws that secure the radiator
to the analyzer, and place the radiator in its storage container.
The analyzers, GC/MS interface, and other components in the analyzer chamber
operate at very high temperatures. Do not touch any part until you are sure it is cool.
• Always wear clean gloves to prevent contamination when working in the analyzer
chamber.
• Make sure you use an antistatic wrist strap and take other antistatic precautions
before touching analyzer components.
• When disconnecting leads, pull on the connectors, not on the wires