7 CI Maintenance
200 5977B Series MSD Operating Manual
To Remove the EI HES Radiator
Materials needed
Gloves, clean, lint-free
Large (8650-0030)
Small (8650-0029)
Tweezers (8710-2460)
T10 Torx screwdriver (5182-3466)
Procedure
1 Remove the EI HES. (See “To Remove the EI HES” on page 154.)
2 Disconnect the wires from the EI HES radiator. (See “To
Connect/Disconnect Wiring to the EI HES Radiator” on page 201.)
3 Use a T10 Torx screwdriver to loosen the 2 screws that secure the radiator
to the analyzer, and place the radiator in its storage container.
WARNING
The analyzers, GC/MS interface, and other components in the analyzer chamber
operate at very high temperatures. Do not touch any part until you are sure it is cool.
CAUTION
Always wear clean gloves to prevent contamination when working in the analyzer
chamber.
Make sure you use an antistatic wrist strap and take other antistatic precautions
before touching analyzer components.
When disconnecting leads, pull on the connectors, not on the wires

Table of Contents

Other manuals for Agilent Technologies 5977B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5977B Series and is the answer not in the manual?

Agilent Technologies 5977B Series Specifications

General IconGeneral
BrandAgilent Technologies
Model5977B Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals