Installing 7890 GC Columns 2
5977B Series MSD Operating Manual 47
To Install a Capillary Column in the GC/MS Interface Using a
Standard Column Nut
This procedure is for the installation of a capillary column directly into the
analyzer. There are two types of column nuts that can be used in the GC/MS
interface: The standard column nut explained here, and the self tightening
column nut explained in “To Install a Capillary Column in the GC/MS
Interface Using the Self-Tightening Column Nut” on page 42.
Materials needed
Interface tip seal (G3870-20542) (See Figure 5 on page 44 and Figure 6 on
page 44.) (Not used with EI SS or Inert source)
Tip seal spring (G7005-20024) (Not used with SS or Inert EI source)
Column cutter, ceramic (5181-8836) or diamond (5183-4620)
Flashlight
Magnifying loupe
Gloves, clean
Large (8650-0030)
Small (8650-0029)
Interface column nut (05988-20066)
Ferrules
0.3 mm id, for 0.10 mm id columns (5062-3507)
0.4 mm id, for 0.20 and 0.25 mm id columns (5062-3508)
0.5 mm id, for 0.32 mm id columns (5062-3506)
0.8 mm id, for 0.53 mm id columns (5062-3512)
Septum (may be old, used inlet septum)
Safety glasses
Wrench, open-end, 1/4-inch and 5/16-inch (8710-0510)