EasyManuals Logo

Agilent Technologies 5977B Series User Manual

Agilent Technologies 5977B Series
228 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #47 background imageLoading...
Page #47 background image
Installing 7890 GC Columns 2
5977B Series MSD Operating Manual 47
To Install a Capillary Column in the GC/MS Interface Using a
Standard Column Nut
This procedure is for the installation of a capillary column directly into the
analyzer. There are two types of column nuts that can be used in the GC/MS
interface: The standard column nut explained here, and the self tightening
column nut explained in “To Install a Capillary Column in the GC/MS
Interface Using the Self-Tightening Column Nut” on page 42.
Materials needed
Interface tip seal (G3870-20542) (See Figure 5 on page 44 and Figure 6 on
page 44.) (Not used with EI SS or Inert source)
Tip seal spring (G7005-20024) (Not used with SS or Inert EI source)
Column cutter, ceramic (5181-8836) or diamond (5183-4620)
Flashlight
Magnifying loupe
Gloves, clean
Large (8650-0030)
Small (8650-0029)
Interface column nut (05988-20066)
Ferrules
0.3 mm id, for 0.10 mm id columns (5062-3507)
0.4 mm id, for 0.20 and 0.25 mm id columns (5062-3508)
0.5 mm id, for 0.32 mm id columns (5062-3506)
0.8 mm id, for 0.53 mm id columns (5062-3512)
Septum (may be old, used inlet septum)
Safety glasses
Wrench, open-end, 1/4-inch and 5/16-inch (8710-0510)

Table of Contents

Other manuals for Agilent Technologies 5977B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5977B Series and is the answer not in the manual?

Agilent Technologies 5977B Series Specifications

General IconGeneral
BrandAgilent Technologies
Model5977B Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals