2 Installing 7890 GC Columns
46 5977B Series MSD Operating Manual
15 Gently check the alignment of the ion source and the interface tip seal.
When the ion source is aligned correctly, the analyzer chamber can be
closed all the way with no resistance except the spring tension from the
interface tip seal.
16 You can align the ion source and interface tip seal by wiggling the side plate
on its hinge. If the door still will not close, contact your Agilent
Technologies service representative.
17 Close the analyzer chamber door. (See “To Close the Analyzer Chamber” on
page 196.)
CAUTION
Forcing the analyzer door closed if these parts are misaligned will damage the tip seal
or the interface or the ion source, or will keep the sideplate from sealing.