6 General Maintenance
170 5977B Series MSD Operating Manual
To Install the EI HES Filament
Materials needed
Filament assembly, high efficiency dual filament (G7001-60061)
Hex ball driver, 1.5 mm (8710-1570)
Gloves, clean, lint-free
Large (8650-0030)
Small (8650-0029)
T6 Torx driver, (8710-2548)
Procedure
1 Remove the old filament. (See “To Remove the EI HES filaments” on
page 168.)
2 Insert the 3 pins on the dual filament through the rear of the filament block.
(See Figure 54 on page 169.)
3 Place the filament block in the source mount.
4 Use a T6 Torx screwdriver to tighten the screw securing the filament block
to the source mount.
5 Reinstall the ion source. (See “To Install the EI HES” on page 171.)
6 Close the analyzer chamber. (See “To Close the Analyzer Chamber” on
page 196.)
7 Pumpdown the MSD. (See “To Pumpdown the MSD” on page 116.)
8 Autotune the MSD.

Table of Contents

Other manuals for Agilent Technologies 5977B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5977B Series and is the answer not in the manual?

Agilent Technologies 5977B Series Specifications

General IconGeneral
BrandAgilent Technologies
Model5977B Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals