7 CI Maintenance
208 5977B Series MSD Operating Manual
To Install the EI HES Radiator
Materials needed
Gloves, clean, lint-free
Large (8650-0030)
Small (8650-0029)
T10 Torx screwdriver (5182-3466)
Tweezers (8710-2460)
Procedure
1 Place the radiator over the guide pins on the analyzer support and use a
T10 Torx screwdriver to secure it using the (2) M3 x 12 gold plated screws
(part number G7002-20110).
2 Use tweezers to disconnect the wires from the EI HES radiator. (See “To
Connect/Disconnect Wiring to the EI HES Radiator” on page 201.)
WARNING
The analyzers, GC/MS interface, and other components in the analyzer chamber
operate at very high temperatures. Do not touch any part until you are sure it is cool.
CAUTION
Always wear clean gloves to prevent contamination when working in the analyzer
chamber.
CAUTION
Make sure you use an antistatic wrist strap and take other antistatic precautions
before touching analyzer components.