7 CI Maintenance
214 5977B Series MSD Operating Manual
To Disassemble the CI Source
Materials needed
Gloves, clean, lint-free
Large (8650-0030)
Small (8650-0029)
Hex ball driver, 1.5 mm (8710-1570)
Hex ball driver, 2.0 mm (8710-1804)
Wrench, open-end, 10 mm (8710-2353)
Nut driver, 5.5 mm (8710-1220)
Tweezers (8710-2460)
Procedure
Refer to the exploded parts view Figure 70 and CI source parts list Table 23 on
page 215 while using this procedure.
1 Remove the CI source. (See “To Remove the CI Source” on page 211.)
2 Remove the filaments. (See “To Remove the CI Source Filamenton
page 223.)
3 Separate the source heater assembly from the source body. The source
heater assembly includes the source heater, repeller, and related parts.
4 Disassemble the repeller assembly by removing the ceramic insulator from
the repeller.
5 Remove the setscrew securing the lenses to the source body.
6 Pull the lenses out of the source body and separate the lens insulator, ion
focus lens, drawout cylinder, drawout lens, and entrance lens.

Table of Contents

Other manuals for Agilent Technologies 5977B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5977B Series and is the answer not in the manual?

Agilent Technologies 5977B Series Specifications

General IconGeneral
BrandAgilent Technologies
Model5977B Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals