1 Introduction
32 5977B Series MSD Operating Manual
To Replace the Primary Fuses
Materials needed
Fuse, T12.5A, 250 V (2110-1398) – 2 required
Screwdriver, flat-blade (8730-0002)
The most likely cause of primary fuse failure is a problem with the foreline
pump. If the primary fuses in your MSD fail, check the foreline pump.
Procedure
1 Vent the MSD, and unplug the power cord from the electrical outlet.
If one of the primary fuses has failed, the MSD will already be off, but for
safety, switch off the MSD and unplug the power cord. It is not necessary to
allow air into the analyzer chamber.
2 Turn one of the fuse holders counterclockwise until it pops out. The fuse
holders are spring loaded. (See Figure 2 on page 33.)
3 Remove the old fuse from the fuse holder.
4 Install a new fuse in the fuse holder.
5 Reinstall the fuse holder.
WARNING
Never replace the primary fuses while the MSD is connected to a power source.
WARNING
If you are using hydrogen as a carrier gas or for the JetClean system, the hydrogen
gas flow must be off before turning off the MSD power. If the foreline pump is off,
hydrogen will accumulate in the MSD and an explosion may occur. Read “Hydrogen
Safety” on page 23 before operating the MSD with hydrogen gas.

Table of Contents

Other manuals for Agilent Technologies 5977B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5977B Series and is the answer not in the manual?

Agilent Technologies 5977B Series Specifications

General IconGeneral
BrandAgilent Technologies
Model5977B Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals