EasyManua.ls Logo

Agilent Technologies 5977B Series

Agilent Technologies 5977B Series
228 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4Operating in EI Mode
106 5977B Series MSD Operating Manual
To Set the Operation Mode for the Optional JetClean System
Procedure:
1 In Instrument Control view, click the MS Parameters icon.
2 In the Single Quadrupole MS Method Editor navigation panel, select the
JetClean tab. (See Figure 33.)
3 From the Operation dropdown menu, select a mode.
For detailed information on the JetClean system see the JetClean Operating
manual.
Figure 33 Selecting the JetClean Operation mode

Table of Contents

Other manuals for Agilent Technologies 5977B Series

Related product manuals