Operating in EI Mode 4
5977B Series MSD Operating Manual 107
To Set the JetClean Parameters for the Clean Only Mode
Procedure:
1 In Instrument Control view, click the MS Parameters icon.
2 In the Single Quadrupole MS Method Editor, select the JetClean tab. (See
Figure 34.)
3 From the Operation dropdown menu, select Clean Only.
4 Enter the parameters supplied by your method developer.
5 Save the method.
Figure 34 JetClean set to Clean Only mode