EasyManua.ls Logo

Agilent Technologies 5977B Series

Agilent Technologies 5977B Series
228 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5 Operating in CI Mode
120 5977B Series MSD Operating Manual
General Guidelines
Always use the highest purity methane (and other reagent gases, if
applicable). Methane must be at least 99.9995% pure.
Always verify the MSD is performing well in EI mode before switching to CI.
(See “To Verify EI System Performance” on page 108.)
Ensure the CI source and GC/MSD interface tip seal are installed.
Ensure the reagent gas plumbing has no air leaks. This is determined in PCI
mode, checking for m/z 32 after the methane pretune.
Ensure the reagent gas inlet line is equipped with gas purifiers (not
applicable for ammonia).

Table of Contents

Other manuals for Agilent Technologies 5977B Series

Related product manuals