EasyManua.ls Logo

Agilent Technologies 5977B Series

Agilent Technologies 5977B Series
228 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Introduction 1
5977B Series MSD Operating Manual 27
If a power failure occurs while the GC/MSD system is unattended, even if
the system has restarted by itself:
1 Immediately turn off the hydrogen at its source.
2 Turn off the GC.
3 Turn off the MSD, and allow it to cool for 1 hour.
4 Eliminate all potential sources of ignition in the room.
5 Open the vacuum manifold of the MSD to atmosphere.
6 Wait at least 10 minutes to allow any hydrogen to dissipate.
7 Start up the GC and MSD as normal.
When using hydrogencheck the system for leaks to prevent possible fire and
explosion hazards based on local Environmental Health and Safety (EHS)
requirements. Always check for leaks after changing a tank or servicing the
gas lines. Ensure the foreline pump exhaust and GC injection port vents are
both vented into a fume hood.

Table of Contents

Other manuals for Agilent Technologies 5977B Series

Related product manuals