Intuvo 9000 GC Column Maintenance 3
5977B Series MSD Operating Manual 67
2 From MassHunter Data Acquisition, Vent the MS (See “To Vent the MSD”
on page 113.)
3 Wait until the GC reaches the ready state indicating the components are
cooled below 40 °C and the instrument is vented before continuing the
steps in this procedure.
4 If you are using hydrogen or other flammable gas as a carrier gas or for the
JetClean system, close the manual gas supply valve to the instrument
before turning off the MS power.
5 If installed, remove the ALS injector from the inlet.
6 Remove the inlet cover.
7 Open the GC front door.
8 Open the bus door and remove it by lifting the door vertically off its hinge
pins.
9 Pull the compression bolt access plate out to allow the torque driver to
access the Guard chip compression bolt.
Figure 13 GC inlet cover and compression bolt access plate
GC inlet cover removed
Compression bolt access plate

Table of Contents

Other manuals for Agilent Technologies 5977B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5977B Series and is the answer not in the manual?

Agilent Technologies 5977B Series Specifications

General IconGeneral
BrandAgilent Technologies
Model5977B Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals