Other Setup Tasks 5
To confirm the collision cell gas flow rate (6470A and 6495A/B only)
Agilent 6400 Series Triple Quad LC/MS System Installation Guide 103
Figure 56 Ion abundance ratio data points as a function of DAC
e Click Set to make permanent the CC DAC value that appears for New
flow rate (DAC).
15 Verify the DAC value:
a In Lab Advisor, open the Backdoor utility.
b Select getccgasflow and press Enter.
c Check the CC gas flow values.