2 Software Installation and Configuration
Step 13. Confirm the LC firmware revision
46 Agilent 6400 Series Triple Quad LC/MS System Installation Guide
Step 13. Confirm the LC firmware revision
1 Start the Data Acquisition program.
2 Click File > Print > Instrument Configuration Report.
3 Click Screen.
4 Click OK.
5 Go to
http://www.chem.agilent.com/_layouts/agilent/downloadFirmware.aspx?w
hid=69761 and check that you have the most recent firmware for your
instrument.
6 If you need to update your LC firmware, follow the instructions on the web
site. Make sure you use the latest firmware update tool, as instructed on the
web site.
7 Check that the firmware for the G1369A Agilent LAN Interface card is the
latest version available:
a Start a Command Prompt in Windows.
b Type telnet 192.168.254.11 in the Command Prompt window and
press Enter.
c Type / in the Telnet window.
The LAN Interface card displays its current configuration information.
The LAN Interface card requires 1.10 firmware or higher. If an upgrade is
needed, please contact your Agilent representative.