EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>6490

Agilent Technologies 6490 Installation Guide

Agilent Technologies 6490
112 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #109 background imageLoading...
Page #109 background image
Reference 6
Negative Ion Mode Checkout Conditions
Agilent 6400 Series Triple Quad LC/MS System Installation Guide 109
Negative Ion Mode Checkout Conditions
These checkout conditions are used in Negative Ion Mode checkout.
Tabl e 1 5 HPLC Pump
Time %A (water) %B (Methanol) Flow
Initial90100.5 mL/min
0.190100.5 mL/min
12980.5 mL/min
22980.5 mL/min
2.01 90 10 0.5 mL/min
3.7 Stop
Tabl e 1 6 HPLC Autosampler
Injection volume 2 µL
Autosampler temperature 4°C
Needle wash flushport 75:25 methanol:water
Needle wash location Flushport, 10 seconds
Tabl e 1 7 Source
ESI (6420/
6430/6460)
AJS (6460) AJS (6470) AJS (6490) AJS (6495)
Gas Temp 310°C 200°C 200°C 210°C 210°C
Gas Flow 13 L/min 6 L/min 6 L/min 17 L/min 17 L/min
Nebulizer 25 psi 35 psi 35 psi 25 psi 25 psi
Sheath Gas Temp 400°C 400°C 400°C 400°C
Sheath Gas Flow 12 L/min 12 L/min 12 L/min 12 L/min
Nozzle Voltage 1000 V 1000 V 2000 V 2000 V

Table of Contents

Other manuals for Agilent Technologies 6490

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6490 and is the answer not in the manual?

Agilent Technologies 6490 Specifications

General IconGeneral
BrandAgilent Technologies
Model6490
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals