Instrument Installation 1
Step 3. Set up the Agilent MS40+ pumps
Agilent 6400 Series Triple Quad LC/MS System Installation Guide 17
11 For 6490/6495: Plug the two pump power cord (p/n 8121-1753) into the
power inlets of the foreline pumps. Secure the pump power cords in place
using the retaining bale mechanisms. Plug the other ends into the Pump
Expander assembly ports. Connect the Ion Funnel Foreline Pump to the
port labeled Auxiliary and connect the Foreline Pump to the port labeled
Foreline. See Figure 9.
12 For 6490/6495: Plug the two pump power cord (p/n 8121-1753) into the
power inlets of the foreline pumps. Secure the pump power cords in place
using the retaining bale mechanisms. Plug the other ends into the Pump
Expander assembly ports. Connect the Ion Funnel Foreline Pump to the
port labeled Auxiliary and connect the Foreline Pump to the port labeled
Foreline. See Figure 9.