1 Instrument Installation
Step 5. Install the spray chamber
28 Agilent 6400 Series Triple Quad LC/MS System Installation Guide
4 Put the nebulizer in the nebulizer adjustment fixture supplied in the
shipping kit and check that the nebulizer needle is properly adjusted. The
nebulizer needle should be even with the end of the nebulizer nozzle.
5 Install the nebulizer in the spray chamber.
6 Install the spray chamber on the spray chamber mount, close the spray
chamber, and fasten the latch. You may need to adjust the latch to ensure
that the O-ring seals completely. Use a ¼-inch x 5/16-inch wrench to loosen
the lock nut, and then adjust the latch to the proper fit, and then tighten the
lock nut so that the latch maintains its adjustment.
7 Connect the high voltage and vaporizer cables to the connectors on the left
side panel of the instrument.
8 Connect the 1/8-inch nebulizing gas tubing from the Triple Quadrupole
LC/MS mainframe to the nebulizer gas fitting.
9 Connect the 1/8-inch sheath gas tubing from the Triple Quadrupole LC/MS
mainframe to the sheath gas fitting on the AJS ESI source.
10 Connect the PEEK tubing from the selection valve (inside the front cover) to
the top of the nebulizer.

Table of Contents

Other manuals for Agilent Technologies 6490

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6490 and is the answer not in the manual?

Agilent Technologies 6490 Specifications

General IconGeneral
BrandAgilent Technologies
Model6490
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals