EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>6490

Agilent Technologies 6490 Installation Guide

Agilent Technologies 6490
112 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #59 background imageLoading...
Page #59 background image
Installation Completion and System Verification 3
Step 3. Prepare the performance evaluation samples
Agilent 6400 Series Triple Quad LC/MS System Installation Guide 59
3 Prepare the second dilution (500 fg/μL) chloramphenicol:
a Use a clean graduated pipette to transfer 1 mL of the 100 pg/μL
chloramphenicol dilution to a pre-rinsed 100 mL volumetric flask.
b Dilute to the 100 mL mark with the 70:30 water:methanol solution.
c Use 5 to 10 mL of this solution to rinse one of the plastic storage bottles
and fill it with the remaining solution.
d Label this bottle “500 fg/μL chloramphenicol” along with the preparation
date.
This provides 500 fg/μL at 2 μL injection volume, which equals the final
1 pg chloramphenicol amount that is needed for performance verification.
Positive Mode (for all 6400 Series LC/MS)
Required Parts 1 mL graduated pipette (p/n 9301-1423)
2 × 100 mL volumetric flasks (p/n 9301-1344)
ES/APCI Positive Ion Performance Standard (p/n G2423A)
Plastic bottle for storing first dilution (p/n 9301-1433)
1 Prepare 2 liters of a 70:30 mixture of acetonitrile:water.
2 Prepare the first dilution (50 pg/μL) reserpine:
a Use a clean graduated pipette to transfer 1.0 mL of the 5 ng/μL reserpine
(p/n G2423A) standard to a pre-rinsed 100 mL volumetric flask.
b Dilute to the 100 mL mark with the 70:30 mixture of acetonitrile:water
solution that was prepared in step 1. Invert the flask at least 10 times to
ensure proper mixing of the checkout sample.
c Use 5 to 10 mL of this solution to rinse one of the plastic storage bottles
and fill it with the remaining solution.
d Label this bottle “50 pg/μL reserpine” along with the preparation date.
3 Prepare the second dilution (500 fg/μL) reserpine:
a a Use a clean graduated pipette to transfer 1 mL of the 100 pg/μL
reserpine dilution to a pre-rinsed 100 mL volumetric flask.
b Dilute to the 100 mL mark with the 70:30 mixture of acetonitrile:water
solution.
c Use 5 to 10 mL of this solution to rinse one of the plastic storage bottles
and fill it with the remaining solution.

Table of Contents

Other manuals for Agilent Technologies 6490

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6490 and is the answer not in the manual?

Agilent Technologies 6490 Specifications

General IconGeneral
BrandAgilent Technologies
Model6490
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals